Sophisticated solution for accurate alignment tasks.
New hardware configuration: a
portable, fast and accurate measurement system having USB2.0
interface. The system has a new hardware with 24 bit A/D, power input
via USB port.
Complete system, including USB manifold box, a measuring head composed
of two Lateral Effect PSD's assembled to a calibrated beam splitter
assembly, including: a beam splitter, a collimated lens and a lens
tube, and a software package for Windows.
Supported: Windows XP, Vista, 7 (32 & 64-bit)
A variety of sensor heads which differ from each other mainly by the
lens focal length. The standard configuration is f=100mm, other common
combinations is f=50mm.
| Download Brochure |
|
|
|
|
|
|
Models:
|
|
ALIGN-USB-50:
ALIGN-USB-100: |
Measuring head with focal length f=50mm
Measuring head with focal length f=100mm (standard configuration)
|
|

|
Models:
|
|
ALIGNLL-USB-50:
ALIGNLL-USB-100:
ALIGNLL-USB-200: |
Measuring head with focal length f=50mm
Measuring head with focal length f=100mm (standard configuration)
Measuring head with focal length f=200mm
|
|

|
Simultaneous measurements of centration and angular deviations with
a large aperture.
Measured parameters are: Position (µm), Angle (mRad), Power (mW).
Complete system composed of a plug-in USB 2.0 interface, a manifold box for
connection of the detector head, PnP detector head assembly (including a PSD, a
lens and lens tube) attached to 3m long cable, software package for Windows,
carrying case, user manual.
Usable beam size up to 22.5mm, angle measurement range +/-2
degrees with angle resolution of 1 sec.
OS supported: Windows 2000/ XP/Vista.
|

|
|
Models:
|
|
ALIGN-LA:
|
Complete system including a measuring head, manifold box with USB 2.0
cable, software on CD disk, carrying case.
|
|
|
|

|
A compact and powerful solution
for laser angular deviations measurements
Measured parameters are: Angle (mRad), Power (mW). Complete system
composed of a measuring head composed of PSD/CCD/CMOS detector plus
lens , a manifold box for connection of detector, 3m long cable
attached to head, software package for Windows, carrying case, user
manual.
Using different focal lengths lens, the system offers large fields
of view with high accuracy, or smaller fields of view with extreme
accuracy. Focal length lens available: 50mm, 85mm, 100mm and 200mm.
Main Features:
- Real time measurement of laser angular displacement and
power.
- Flexible configurations.
- CW or pulsed beams.
- Multiple device control via USB.
- TCP/IP communication protocol and remote control.
- ActiveX software for integration in customer application
program.
OS supported: Windows 98 / ME / 2000 / XP.
|

|
|
|
|
|
|
Models
|
CMOS USB2
|
CCD USB2
|
PSD4
USB
|
PSD4
USB HR
|
PSD9
USB
|
PSD4 Analog
|
PSD4 Analog HR
|
PSD9 Analog
|
|
Optics
Focal lens
|
50mm f1.8
|
50mm f1.8
|
50mm f1.8
|
50mm f1.8
|
50mm f1.8
|
50mm f1.8
|
50mm f1.8
|
50mm f1.8
|
|
Sensor
Size
(mm)
|
4.7 x 3.4
|
4.6 x 6.2
|
4 x 4
|
4 x 4 Work Area 1x1
|
9 x 9
|
4 x 4
|
4 x 4 Work Area 1x1
|
9 x 9
|
|
Field
of
View
|
90X65 mRad
|
120X90 mRad
|
80X80 mRad
|
20X20 mRad
|
180X180 mRad
|
80X80 mRad
|
20X20 mRad
|
180X180 mRad
|
|
Angle
Resolution
|
20µRad
|
20µRad
|
15µRad
|
3µRad
|
30µRad
|
5µRad
|
1µRad
|
10µRad
|
|
Angle
Accuracy
|
Max dev. 100µR
|
Max dev 100µR
|
0.5%
|
0.5%
|
0.5%
|
0.5%
|
0.5%
|
0.5%
|
|
Power
Measurement
|
No
|
No
|
Yes
|
Yes
|
Yes
|
Yes
|
Yes
|
Yes
|
|
Power
accuracy
|
---
|
---
|
±5%
|
±5%
|
±5%
|
±5%
|
±5%
|
±5%
|
|
Interface
|
USB2 Camera
|
USB2 Attach.
|
Analog &USB1
|
Analog &USB1
|
Analog &USB1
|
Analog 0-10V
|
Analog 0-10V
|
Analog 0-10V
|
| Update rate
|
30Hz
|
30Hz
|
Analog 60KHz
Digital 40Hz
Burst 4KHz
|
Analog 60KHz
Digital 40Hz
Burst 4KHz
|
Analog 30KHz
Digital 40Hz
Burst 4KHz
|
60KHz
|
60KHz
|
30KHz
|
|
Response
time
|
nSec
|
nSec
|
>20µS
|
>20µS
|
>60µS
|
>20µS
|
>20µS
|
<60µS
|
|
Multiple
beams meas.
|
3 beams simultan.
|
3 beams simultan.
|
Several units in parallel
|
Several units in parallel
|
Several units in parallel
|
No
|
No
|
No
|
|
 |
A compact and powerful solution
for laser angular deviations measurements
Measured parameters are: Angle (mRad), Power (mW).
Complete system composed of a plug-in card (ISA / PCI / PCMCIA), a manifold box
for connection of detector/s, PnP detector head assembly (PSD plus lens and
lens tube) attached to 3m long cable, software package for Windows, carrying
case, user manual.
Standard configuration is 100mm F.L.
OS supported: Windows 95 / 98 / NT 4.0 / ME / 2000 (Excluding NT 4.0 for the PCMCIA version).
|

|
|
Models:
|
|
Angle-ISA:
Angle-PCI:
Angle-PCMCIA:
Angle-ANALOG:
|
Complete system based on an ISA card
Complete system based on a PCI card
Complete system based on a PCMCIA card
Complete system based on an analog electronics, AngleMeter
measuring head with 3m long attached cable.
|
|
|